AIR SENTRY FUME HOOD
INDEPENDENT PERFORMANCE TEST REPORTS

The files below are pdf files.  

Need Adobe Reader
click here

 
 


 

 

ECT-HBASC6.PDF 
Exposure Control Technologies ASHRAE 110 testing
on a 6' bench top Air Sentry Fume Hood at Lab Crafters facility on 9/13/99.

ECT-HWASC6.PDF Exposure Control Technologies ASHRAE 110 testing on a 6' walk-in Air Sentry Fume Hood at Lab Crafters facility on 9/15/99.

UWM-HBASC5.PDF Modified ASHRAE 110 ("As Installed" and simulated "As Used") testing on a 5' bench top Air Sentry Fume Hood at the University of Wisconsin-Milwaukee on 2/21/00.

UWM-HBASC5-SUM.PDF Same report as above, no appendices.

The Air Sentry Fume Hood's performance has been tested to every nationally and internationally recognized testing standard in the world, including the ASHRAE 110-1995 Standard, the BS 7258: Part 4: 1994 Standard, The Invent UK Test Method, and the DIN 12 924 Standard.

Lab Crafters has further challenged the performance of its hoods with tests that go above and beyond the scope of these standards. These additional challenges include greater tracer gas release rates, varying and multiple tracer gas sensor locations, and the addition of cross drafts and other dynamic challenges during the tests.

The test reports above are three of the many performance tests conducted on the Air Sentry Fume Hood by independent organizations. To download the reports in *.pdf format, simply click on the bold type file name. Additional test reports will be posted as they become available in digital format.

The first two reports (ECT-HBASC6.PDF and ECT-HWASC6.PDF) were performed by Mr. Thomas C. Smith of Exposure Control Technologies, of Cary, North Carolina. Mr. Smith is recognized as one of the top experts in the field of fume hood testing and commissioning and is among the most experienced individuals in the world in the ASHRAE 110 test method. These reports contain the results of ASHRAE 110-1995 tests performed on a six-foot wide bench top Air Sentry Fume Hood and a six-foot wide walk-in Air Sentry fume hood, respectively. These tests were conducted at Lab Crafters test facility in Ronkonkoma, NY.

The third report (UWM-HBASC5.PDF) was prepared by the State of Wisconsin, Department of Administration, Division of Facilities Development. This report contains the results of modified ASHRAE 110-1995 tests, performed by Mr. Smith of ECT and Mr. Jack Wunder of the University of Wisconsin-Madison, on a five-foot wide bench top Air Sentry Fume Hood. For this report, the Air Sentry Fume Hood was tested side by side with a Fisher Hamilton, Inc. HOPEC Fume Hood at the University of Wisconsin-Milwaukee. The report compares the performance of the two hoods.

Any questions regarding these test reports may be forwarded to Lab Crafters or to the organizations that prepared the reports.

 