Proven Test Results

When tested to the ASHRAE 110-1995 Method of Testing Laboratory Fume Hoods, the Air SENTRY performs exceptionally well. The Air SENTRY qualifies as a "Class A" fume hood, per the performance specification in the ANSI/AIHA Z9.5-1992 American National Standard for Laboratory Ventilation. These ASHRAE tracer gas test results show a standard 6' bench Air SENTRY'S performance, with a face velocity of 40 FPM, compared to the performance of two competitors, standard 6' bench hoods, with face velocities of 100 FPM. ASHRAE "As Manufactured" test reports are available upon request.

 

INDEPENDENT PERFORMANCE TEST REPORTS

FOR THE
AIR SENTRY FUME HOOD
click here for details

 

Dynamic "Space Pressurization Effect" Fume Hood Performance Testing

 

Dynamic "Sash Movement Effect" Fume Hood Performance Testing

 

 

 

 

 

AIR SENTRY AIRFLOW REQUIREMENTS

Hood
Size

Exhaust
Volume
CFM

Sash Opening

Face Velocity FPM

Static
Pressure
Collar Size
Vertical Horizontal Vertical Horizontal
4'
4'
4'
4'
220
300
340

340
20"x39"
28"x39"
25"x39"
21"x39"
26.5"x18.5"
26.5"x18.5"
26.5"x18.5"
26.5"x18.5"
40
40
50
60
65
88
100
100
.10"wg
.19"wg
.22"wg
.22"wg
8"
8"
8"
8"
5'
5'
5'
5'
270
395
450
450
19"x51"
28"x51"
25"x51"
21"x51"
26.5"x24.5"
26.5"x24.5"
26.5"x24.5"
26.5"x24.5"
40
40
50
60
60
87
100
100
.17"wg
.33"wg
.17"wg
.17"wg
8"
10"
10"
10"
6'
6'
6'
6'
330
490
560
560
19"x63"
28"x63"
25.5"x63"
21"x63"
26.5"x30.5"
26.5"x30.5"
26.5"x30.5"
26.5"x30.5"
40
40
50
60
60
87
100
100
.22"wg
.18"wg
.27"wg
.27"wg
8"
10"
10"
10"
8'
8'
8'
8'
470
600
780
780
19"x87"
24.5"x87"
25.5"x87"
21.5"x87"
26.5"x42.5"
26.5"x42.5"
26.5"x42.5"
26.5"x42.5"
40
40
50
60
60
77
100
100
.14"wg
.19"wg
.33"wg
.33"wg
(2) 8"
(2) 8"
(2) 8"
(2) 8"

 

Notes: The lowest exhaust volumes for each size fume hood are per the minimum exhaust volume recommendations stated in NFPA 45, Section A-6-4.5. The static pressure values shown represent the pressure loss for the fume hood only. Inquire for Walk-In style Air Sentry air flow information

 

 

shade.gif (2128 bytes)

Home | Products | Company Bio | Site Map | Employment | Contact Us

 

Lab Crafters, Inc.
2085 Fifth Avenue
Ronkonkoma, NY 11779
(631) 471-7755 Voice
(631) 471-9161 FAX

info@lab-crafters.com
http://www.lab-crafters.com

logosm.gif (3223 bytes)
© 1998 Lab Crafters, Inc., All Rights Reserved.
We welcome your feedback.